A study on how sash movement affects performance of fume hoods.

This study was conducted to determine how sash movements affect the performance of fume hoods. The performance of two fume hoods was studied as the sashes were moved from closed to open position at speeds of 2 ft/s, 1.5 ft/s, and 1 ft/s. The tests were conducted with fume hoods operated at both constant volume and variable air volume. The tests indicate that sash movements can disturb airflow patterns at the face of the hood and potentially affect the performance of the hood. The effect of the sash movement varied with hood type and speed of sash movement.

Laboratory fume hood and exhaust fan penthouse exposure risk analysis using the ANSI/ASHRAE 110-1995 and other tracer gas test methods.

The use of the laboratory fume hood as the primary containment device in the laboratory has been a standard practice for almost half a century. Quantitative testing of the performance of these devices, however; is a more recent discipline. The use of the ANSI/ASHRAE 110-1995, Method of Testing Performance of Laboratory Fume Hoods (ASH RAE 1995) is becoming a standard specification in the purchase of new fume hoods, the commissioning of new laboratory facilities, and benchmarking fume hoods in existing facilities.