Paul Simons, Stefanie Rolfsmeier
Year:
2015
Languages: English | Pages: 10 pp
Bibliographic info:
36th AIVC Conference " Effective ventilation in high performance buildings", Madrid, Spain, 23-24 September 2015.

This presentation will explain what exactly calibration should mean. It will also look into the hierarchy of and the differences between the calibration labs of national standards, accredited labs (like DAkkS, a national government-appointed accreditation body in Germany, or Cofrac – Comité français d’accréditation – in France), and manufacturer’s calibration labs. Even the labs themselves must have their own measuring devices checked in order to guarantee what is known as traceability to national standards, and to determine the measurement uncertainty of their testing devices. This is necessary if you want to make a clear statement on adherence to the accuracy, which is known as a statement of compliance or declaration of conformity. Another topic covers the measuring range and the accuracy of pressure gauges and measuring fans, as well as their calibration process, by looking at the example of the Minneapolis BlowerDoor.