Describes methods of measuring the three-dimensional flow field using a three-sensor hot-wire probe, with emphasis on the techniques developed by the author's group at the Pennsylvania State University. The hot-wire equations, data processing procedure, calibration techniques, and a discussion of various errors in the measurement are included. Some typical data acquired by thisprobe is also included.
Three sensor hot wire/film technique for three dimensional mean and turbulence flow field measurement.
Year:
1982
Bibliographic info:
TSI quarterly vol.8 no.1 January-March 1982 p.3-30